The basics

The MDR (Regulation (EU) 2017/745) has applied since 26 May 2021. An AI medical device must meet the MDR, and it may also be a high-risk AI system under the AI Act. The key guidance is MDCG 2025-6 / AIB 2025-1, a June 2025 FAQ written jointly by the Medical Device Coordination Group and the AI Board. It is not legally binding.

When is an AI medical device high-risk?

Under AI Act Article 6(1), when two conditions are met: the AI is a safety component or is itself a device, and a notified body performs the conformity assessment.

MDR classHigh-risk under the AI Act?
Class I sterile, measuring, reusable surgicalYes
Class IIa, IIb, IIIYes
Most Annex XVI productsYes
Class I non-sterile, non-measuringNo
In-house devices (Art. 5(5))No

The AI Act does not change MDR risk classes.

How compliance is combined

  • One conformity assessment under the MDR; the notified body also checks AI Act requirements.
  • One technical file.
  • AI Act quality management elements can sit inside the MDR quality system.
  • Data governance (Art. 10): representative, bias-checked datasets.
  • Human oversight and transparency (Arts. 13-14), AI-specific cybersecurity, and real-world testing through clinical investigations.
  • Changes planned and assessed in advance are not substantial modifications.

Dates after the 2026 amendment

The FAQ gave 2 August 2027 for Annex I (product-embedded) high-risk duties. The Digital Omnibus on AI (Regulation (EU) 2026/1744) moved this to 2 August 2028, and lets implementing acts limit duplication with sector law such as medical devices. Separately, the Commission’s proposal for a targeted MDR/IVDR revision (COM(2025) 1023, 16 December 2025) is still a proposal.

Reuse terms

Commission health pages are CC BY 4.0 under the Commission legal notice; the FAQ states it is not a Commission document. MDR text on EUR-Lex is reusable with attribution; only the Official Journal is authentic.

Using these sources in AI and RAG

  1. Index the MDR, AI Act, and FAQ as separate layers.
  2. Update the FAQ’s dates with a note on the 2026 amendment.
  3. Tag devices by class to answer the high-risk question.
  4. Map AI Act requirements into your MDR technical file structure.
  5. Keep proposals labeled until adopted.

Common mistakes

  • Assuming all AI medical devices are high-risk.
  • Planning two separate conformity assessments.
  • Using the FAQ’s 2027 date.
  • Treating the MDCG FAQ as binding.

Checklist

  • Device class and notified body involvement checked
  • AI Act requirements in the technical file
  • QMS updated for AI Act elements
  • Dates updated for 2028
  • Layers separated in index

How AskDona helps

AskDona checks your technical file against MDR and AI Act requirements together, citing each article and showing where the evidence sits.

Run a sample MDR and AI Act technical file check

AskDona shows the evidence; your regulatory team decides.

Editor’s note. This article is general information, not legal advice. The “conditions to check” rating comes from our source list and is a first-pass judgment, not a legal opinion.

FAQ

Is my AI medical device high-risk under the AI Act? If it needs notified body assessment under the MDR, it generally is.

Do I need two conformity assessments? No. One MDR assessment covers AI Act requirements too.

When do AI Act duties apply to AI medical devices? From 2 August 2028, after the 2026 amendment.

Is MDCG 2025-6 binding? No, it is guidance.

References

All pages opened and checked by October 26, 2026.

SourceWhat it supports
MDCG 2025-6 / AIB 2025-1 FAQ (PDF)High-risk criteria, combined assessment
Commission: MDR/IVDR overviewMDR dates, COM(2025) 1023
Commission: MDR/IVDR simplification proposalProposal
Council: AI Act final green light (Jun 2026)Annex I date 2 Aug 2028
European Commission: Legal noticeCC BY 4.0