For nearly 30 years, “quality system” in the US meant FDA’s Quality System Regulation (QS regulation, or QSR), 21 CFR Part 820. On February 2, 2026, Part 820 became the Quality Management System Regulation (QMSR), which incorporates ISO 13485:2016 by reference (FDA). FDA published the final rule on February 2, 2024, giving manufacturers two years to transition (Federal Register).

More than eight months after the effective date, many companies have rewritten their quality manual but not their procedures, forms, training or audit checklists. This guide explains what changed, what did not, and how to confirm your transition is actually complete.

QS regulation vs QMSR at a glance

TopicQS regulation (until Feb 1, 2026)QMSR (from Feb 2, 2026)
Core requirementsDetailed FDA text, §§820.20–820.250ISO 13485:2016 incorporated by reference (§820.10(a))
FDA-specific additionsEmbedded throughoutConcentrated in §820.10(b)–(c), §820.35 (records) and §820.45 (labeling and packaging)
Risk managementMainly in design validationRisk-based approach across the QMS per ISO 13485
Design recordsDesign history file (DHF)Design and development file (ISO 13485 clause 7.3.10)
Device-level documentationDevice master record (DMR)Medical device file (ISO 13485 clause 4.2.3)
Production recordsDevice history record (DHR)Records required by ISO 13485, including clause 7.5.1 and 7.5.8 traceability
Audit and review recordsShielded from routine FDA review by §820.180(c)Open to inspection: exception “not maintained”
Inspection approachQSIT (four subsystems)CP 7382.850 (six QMS areas)

Sources: eCFR Part 820; FDA QMSR FAQ; FDA CP 7382.850.

What changed in practice

ISO 13485 became the requirement text. §820.10(a) requires manufacturers to “Document a quality management system that complies with the applicable requirements of ISO 13485” (eCFR). Procedures should now cite ISO 13485 clauses.

FDA-specific requirements were consolidated, not dropped. §820.10(b) ties the QMS to UDI (Part 830), tracking (Part 821), medical device reporting (Part 803) and corrections and removals (Part 806). §820.35 adds specific content to complaint and servicing records, and §820.45 keeps detailed labeling and packaging controls (eCFR).

More records are visible. FDA states that “The exceptions that existed in the QS regulation at § 820.180(c) are not maintained” (FDA QMSR FAQ). Management review, internal audit and supplier audit reports can be reviewed during inspections.

Inspections changed. QSIT was retired. CP 7382.850 inspects six areas: change control; design and development; management oversight; measurement, analysis and improvement; outsourcing and purchasing; and production and service provision (FDA).

Risk moved to the center. Early QMSR inspections show “Risk management is the top observation” (RAPS).

What did not change

  • FDA still does not certify quality systems. An ISO 13485 certificate does not replace FDA inspection, and FDA inspections do not follow certification audit methods.
  • Other FDA regulations still apply in full: registration and listing, premarket requirements, labeling, MDR, corrections and removals, UDI and Part 11 for electronic records.
  • Design controls remain mandatory for Class II and III devices and specified Class I devices, now via ISO 13485 clause 7.3 (§820.10(c)).
  • Process validation, CAPA, complaint handling and supplier controls are still core; the requirement text moved, the expectations did not soften.

Legacy QSR to ISO 13485 cross-reference

Use this map to find procedures that still cite retired sections. It is a practical starting point, not a one-to-one legal equivalence; some QSR requirements are spread across several ISO 13485 clauses.

Legacy QSR sectionTopicPrimary ISO 13485:2016 clause(s)
820.20Management responsibility5.1–5.6
820.22Quality audit8.2.4
820.25Personnel and training6.2
820.30Design controls7.3 (required by §820.10(c))
820.40Document controls4.2.4
820.50Purchasing controls7.4
820.70 / 820.75Production and process controls; process validation7.5.1, 7.5.6
820.72Inspection, measuring and test equipment7.6
820.80 / 820.86Acceptance activities and status7.4.3, 7.5.8, 8.2.6
820.90Nonconforming product8.3
820.100CAPA8.5.2, 8.5.3
820.120 / 820.130Labeling and packaging7.5.1 plus §820.45
820.180–820.186Records, DMR, DHR, QS record4.2.3, 4.2.5 plus §820.35
820.198Complaint files8.2.2 plus §820.35 and Part 803
820.200Servicing7.5.4 plus §820.35
820.250Statistical techniques8.1, 8.4

Terminology to update

Old term (QSR)QMSR / ISO 13485 term
Management with executive responsibilityTop management
Design history file (DHF)Design and development file
Device master record (DMR)Medical device file (MDF)
Device history record (DHR)Records of each medical device or batch (ISO 13485 clause 7.5.1)
Quality system recordQMS documentation per clause 4.2

FDA’s December 2025 webinar on QMSR key takeaways walks through these changes and is a useful training resource (FDA).

QMSR transition completion checklist

  • Quality manual references ISO 13485:2016 and Part 820 (QMSR), with scope and exclusions justified
  • Every procedure, work instruction and form updated to ISO 13485 clauses; no live references to retired QSR sections
  • DHF, DMR and DHR terminology replaced, or mapped in a controlled glossary
  • §820.10(b) links in place: UDI, tracking, MDR and corrections and removals procedures
  • Complaint and servicing records include §820.35 content
  • Labeling and packaging procedures meet §820.45
  • Risk-based approach documented across QMS processes, not only design
  • Management review, internal audit and supplier audit reports written for FDA visibility
  • Internal audit checklists rebuilt around FDA’s six QMS areas
  • QMS software validated using a risk-based CSA approach (FDA)
  • Training completed on the new structure and terminology
  • Mock inspection run against CP 7382.850

Find the leftovers fast

The hardest part of a QMSR transition is finding what was missed: a form that still cites §820.198, a procedure that never mentions risk, a complaint record missing §820.35 content. AI document assessment can read your full document set against the QMSR and ISO 13485, flag outdated references and missing requirements, and cite the passage behind every finding.

Run a sample regulatory document assessment on a set of procedures. AskDona shows the evidence; your quality team decides.

FAQ

What is the medical device quality system regulation now? The Quality Management System Regulation (QMSR), 21 CFR Part 820, effective February 2, 2026, which incorporates ISO 13485:2016 (FDA).

Do DHF, DMR and DHR still exist? Not as regulatory terms. Their content maps to the design and development file, the medical device file and production records under ISO 13485.

Can FDA now review internal audit reports? Yes. The §820.180(c) exception is “not maintained” under QMSR (FDA QMSR FAQ).

Does ISO 13485 certification mean we comply with QMSR? No. QMSR includes FDA-specific requirements beyond ISO 13485, and FDA inspects independently of certification bodies.

References

All pages opened and checked by October 13, 2026.

SourceTypeDateUsed for
FDA: QS Regulation / QMSRRegulator2026Effective date
Federal Register: QMSR final ruleRegulationFeb 2, 2024Final rule
eCFR: 21 CFR Part 820RegulationCurrent§§820.10, 820.35, 820.45
FDA: QMSR FAQRegulator2026§820.180(c)
FDA: CP 7382.850RegulatorFeb 2, 2026Inspection areas
FDA: QMSR key takeaways webinarRegulatorDec 16, 2025Training
RAPS: Top QMSR observationsProfessional bodyMay 14, 2026Risk findings
FDA: CSA guidanceRegulator guidanceFeb 3, 2026Software validation